Ivchenko L.G. 1
Pervushin Y.V. 1
1 Medical University “Stavropol State Medical University” of the Ministry of Health of the Russian Federation
Assess the possibility of determining the concentration indices of serum lipids on nonchemical acoustic analyzer ACBA -01- «BIOM ®». Conducted parallel investigations control sera with normal and abnormal lipid and serum of healthy donors acoustic method and conventional biochemical methods for automatic biochemical analyzer Kone Delta. Grading is carried out by determining: 1) Intra- reproducibility, 2) Inter-assay reproducibility and accuracy of measurements using control sera, and 3) the definition of parallel lipid metabolism on the unit ACBA -01- «BIOM ®» and automatic biochemical analyzer Kone Delta. It was found that the determination of lipid acoustic method analyzer ACBA-01-«BIOM ®» reproducibility and accuracy does not differ from analyzes performed on the analyzer Kone Delta and fully complies with regulations governing the quality of the definition of these analytes.